White Balance Calibration

Note

  • To ensure that the colors are reproduced accurately, perform the white balance correction every 3 months.

  • White Balance Calibration is only supported by AutoScan-DS-EX, AutoScan-DS-EX Pro(C) and AutoScan-DS-EX Pro(H).

Steps

  1. Click and select White Balance.

  2. Secure the white cover to the calibration board.

  3. Place the height adaptor on the turntable.

  4. Place the assembled calibration board on the height adaptor.

  5. Click Start so that the turntable starts to spin and the white balance correction correction will processed automatically.

  6. Click to confirm and exit.