Calibration notice¶
Through calibration, all the parameters are recalculated to ensure the accuracy of the device and the quality of scan data.
Note
Calibration is required when:
- The scanner is used for the first time.
- The scanned data is incomplete or its quality drops.
- The scanner is severely shaken or jolted, for example, during transport.
- The accuracy drops during scanning, such as frequent alignment failures or frequent indications of tracking lost.
- The scanner has not been calibrated for more than 14 days. In this situation, the software will prompt you to calibrate the scanner.
Caution
- Do not wipe it with any chemical liquid.
- Use the calibration board supplied with the device.
- The calibration board are only used for calibration.
- Please put the calibration board in the flannel bag after use.
- Do not place heavy objects or sundries on the calibration board.
- Always make sure that both sides of the calibration board are clean and free of scratches.
- Keep the calibration board away from corrosives, metals and sharp objects to avoid corrosion or damage.