Quick calibration

Through calibration, the device parameters are recalculated, which not only ensures the accuracy of the device, but also improve the scanning quality.

Note

Calibration is required under the following conditions:

  • When the scanner is used for the first time, or when it is reused after being laid idle for a long period of time (1-2 weeks).

  • The scanner was severely shaken or vibrated, such as shaken during transportation.

  • Severe accuracy reduction, such as frequent errors in alignment or unrecognized markers.

  • Incomplete data is acquired during the scanning or serious deterioration of the quality of scanned data.

Caution

  • Start to calibrate when the temperature of the device get high enough(35℃).

  • The calibration board is matched to the device. Calibrating with an mismatched calibration board will lead to inaccuracy or fail to get good scanned data.

  • Always make sure that both sides of the calibration board are clean and free of scratches.

  • Do not place heavy objects or sundries on the calibration board.

  • Keep the calibration board away from corrosives, metals and sharp objects to avoid corrosion or damage.

  • It is not recommended to wipe the calibration board. If needed, gently wipe it with a piece of a clean damp cloth. Do not use a cloth with chemicals or alcohols to wipe the calibration board.

  • After using the calibration board, put it in a flannel bag.

Step

Follow the steps provided by the calibration wizard on the right side of the interface.

  1. Place the calibration board horizontally.

  2. Place the scanner in the same direction as shown in the figure.

  3. Align the center point of the Device’s handle with the center point of the gray circle on the calibration board.

  4. Press the scan button on the scanner to start calibration.

  5. Move the device slowly and adjust the distance between the scanner and the calibration board according to the height indicating box.

  6. Keep moving until all height boxes turn green.

  7. Check the calibration result.

Note
●If the calibration fails, please try it again.
●If the calibration still fails after several attempts, please contact the supplier or technical support.