Preparation

You can re-adjust parameters of the device through calibration, which not only ensures the accuracy of the device, but also the scanning quality.

Note

Calibration is required under the following conditions:

  • The device was severely shaken or vibrated during transportation.
  • The device is used for the first time or it is not in use for a long period of time (1-2 weeks).
  • The accuracy drops during scanning, such as frequent alignment failures or the inability to recognize markers.
  • Incomplete data is acquired or the quality of scanned data drops during scanning.

Caution

  • Do not wipe the calibration board with chemical liquids.
  • Do not put heavy objects or sundries on the calibration board.
  • After using the calibration board, put it in the flannel bag.
  • The calibration board can only be used for the calibration of the device.

  • Make sure both the scanner and tracker are online before or during calibration.

  • Keep the calibration board away from corrosives, metals and sharp objects to avoid corrosion or damage.

  • Make sure that the markers on the calibration board aren't damaged or stained, and that the front side of the calibration board is clean and free of scratches.
  • After powering on the device, please warm it up to a proper and stable temperature (about 35℃) before the calibration to ensure accuracy and precision.