Cleaning and Sterilization¶
Steps¶
- Rinse: Remove the scanbody from the patient's mouth and take the screw out (lift the screw from the interface of the scanflag with a driver). Then immediately rinse the scanflag and the screw under flowing water for at least 1 minute to preliminarily remove surface contaminants. Hand washing is recommended, with water temperature between 15℃ and 30℃.
- Soak: Submerge the rinsed scanflag and the screw into neutral enzymatic cleaning solution (3M Rapid Multi-Enzyme Cleaner is recommended with a suggested ratio of 5 mL ~ 10 mL of cleaner to 1 L of water). Soak for 2 ~ 10 minutes at a temperature of 25℃ ~ 40℃ to prevent blood coagulation in the screw channel (if the cleaning solution used is not the recommended one, please follow its instructions).
- Clean: During soaking, gently agitate to further clean the screw channels by rotating the liquid until soaking is complete.
- Rinse: Rinse residual enzymatic cleaning solution from the scanflag and the screw with purified water for at least 1 minute. Ensure no contaminants remain on the surface.
- Dry: Gently wipe the scanflag and the screw with a clean, lint-free cloth. Or use an air blow gun to dry the scanflag and the screw. Please ensure that they are dry before sterilization.
-
Sterilization: Seal the screw with an autoclave bag; place the scanflag in the provided sterilization box and seal the box with an autoclave bag; place two bags in the autoclave machine. Sterilization parameters:
- Gravity-displacement sterilization: temperature: 121℃; sterilization time: 30 min; drying time: 30 min.
- Dynamic-air-removal sterilization: temperature: 134℃; sterilization time: 4 min; drying time: 30 min.
Caution:
- One sterilized scanbody can only be used for one person.
- The scanbody should be sterilized. As a consumable item, please replace the scanbody promptly after 300 times of autoclave.
- Do not use metal tweezers to clamp the black part at the top of the scanbody; you can clamp the metal at the bottom to prevent damage to the coding points on the scanbody's surface.
- If obvious scratches are found on the scanbody's surface, stop using it immediately and replace it.
- After drying, please wait until the scanbody cools to room temperature before taking it out to prevent from scalding.
- Do not use ultrasonic wave to clean the scanbody to prevent the coded markers on its surface from being damaged.
- Auto-cleaning is not supported. Please clean the scanbody manually.
- Before sterilization, please wear clean surgical gloves.
Warning:
- There is aluminum in the product. Please avoid using cleaning solution that is sensitive to this material.
- Oxidizing agents and strong alkaline agents, such as hydrogen peroxide and potassium hydroxide, can not be used for cleaning the scanbody.
- Do not use tap water to clean the scanbody, instead, use purified water.