Pre-cleaning, Disinfection, and Sterilization¶
The whole set of scanner, including scanner tip, scanner body, and scanner cradle, requires proper care, cleaning, and handling. As individual part may be processed differently, read and follow the information and instructions given to help you effectively and thoroughly reprocess the set.
Caution
-
All parts are shipped non-sterilized. Follow the instructions prior to initial use.
-
Ensure that you have completely disconnected the power supply and all connections from the scanner.
-
Follow the instructions given in the Manual to pre-clean, disinfect, and sterilize each part of the scanner. Using other methods not approved by SHINING 3D will damage your scanner and void your warranty.
-
Using detergent, disinfection solutions or wipes, sterilization procedures other than those specified in the Manual may damage the product and void your warranty.
-
Only sterilize the part(s) for which a sterilization method is specified. Do not attempt to sterilize all parts of the product. SHINING 3D is not liable for any damages due to improper sterilization.
-
After sterilization, wait until each of the parts is at room temperature to prevent possible heat injuries to the user and the patient.
-
To prevent cross-contamination, pre-cleaning, disinfection, and sterilization must be correctly performed after each use.
-
When the scanner tip is detached from the scanner, always protect the subtle units and the inner optical components on the front end of the scanner body by putting on the supplied dust cap.