Read Before Calibration¶
Read the following content before calibration.
Caution
-
Please do not touch the front side of the calibration board and do ensure that there are no scratches or dirt on the calibration board.
-
Please keep the calibration board away from corrosive liquid, metals, or sharp substances.
-
Don't place heavy objects on the calibration board.
-
Don’t wipe the calibration board with any chemical liquid including alcohol.
-
Please put the calibration board in the storage bag after finishing scanning.
The device needs calibration in the following scenarios:
Note
-
First-time connection;
-
The scanner experienced violent shaking;
-
A deterioration in the quality of the scanned data or frequent errors when stitching compared to the results of previous scans;
-
Changes have occurred in the external environment during use. For example, the ambient temperature varies considerably;
-
Calibrating the device twice a month is recommended.
Calibration may fail in the following scenarios:
Note
-
Calibration without the height adaptor;
-
mismatched calibration files;
-
Excessive ambient brightness. For example, calibrations in bright light;
-
The front of the calibration board is badly damaged.