Operation¶
The entire calibration process includes camera calibration and white balance calibration. You can complete the corresponding calibration according to your needs.
Binding the Calibration Board¶
Before the first calibration or after changing the calibration board, please bind the calibration board first, then start the calibration.
Method 1:
- After connecting the device, click
Calibration in the navigation bar at the top of the interface to switch to the calibration interface. - Click
to the right of the calibration board number in the lower right corner of the interface to open the binding calibration board window. - Click Start Scanning.
- Use the device to scan the QR code on the calibration board to complete the binding.
Note
- If this is your first calibration, the binding calibration board window will automatically pop up after switching to the calibration interface. You can use the device to scan the QR code on the calibration board to complete the binding.
- If the light is too dim, click
or press the middle button on the device to turn on the device's fill light before scanning the QR code.
Method 2:
- Click
>
About > Calibration Board No.
to open the binding calibration board window. - Click Start Scanning.
- Use the device to scan the QR code on the calibration board to complete the binding.
Note
- If the light is too dim, click
or press the middle button on the device to turn on the device's fill light before scanning the QR code.
Camera Calibration¶
If you are using the device for the first time, have not recalibrated for a long time, or have changed the calibration board, please perform camera calibration.

Caution
- Do not perform calibration on reflective tile floors.
- Do not perform calibration in environments with cluttered markers.
Steps
- Please place the calibration board with the marked side facing up on the flat surface as shown.
-
Rotate the scanner to align the white short line
with the red arrow
, then adjust the scanner angle so that the center of the yellow ball aligns with the center of the white circle.

-
Once everything is aligned, a yellow crosshair appears on the screen. Move the crosshair into the circle; the crosshair turns green.

-
With the crosshair green, move the scanner up and down until the blue area on the screen is completely eliminated.

Caution
-
When moving the scanner up and down, please ensure:
- The center of the scanner remains aligned with the center of the calibration board.
- The scanner remains parallel to the plane where the calibration board is located.
-
If you are unable to eliminate the blue area completely while moving the scanner up and down, try adjusting the camera brightness using the brightness slider on the left side.
-
-
Place the calibration board on the holder as shown.
- Repeat steps 2 to 5 to complete the calibration for that direction.
- Rotate the calibration board as shown, and repeat steps 2 to 5 to complete the calibration for other directions.
- After completing all calibration steps, a calibration file will be automatically generated.
- If the calibration is successful, you can exit calibration and proceed to scanning; if the calibration fails, please recalibrate.
Note
- After calibration is complete, please store the calibration board properly for future use.
- If calibration fails multiple times, please promptly contact technical support and provide the error code.
White Balance Calibration¶
White balance calibration is an optional calibration step. It can be performed when there is a discrepancy between the scanned texture data and the actual texture of the object.
Steps
- Please place the calibration board with the white side facing up on a flat surface.
- Aim the scanner at the calibration board.
- Move the scanner vertically up and down until the white circle on the screen completely overlaps with the blue circle.
- After successful calibration, you can exit calibration and proceed to scanning; if calibration fails, please recalibrate.
Note
- After calibration is complete, please store the calibration board properly for future use.
- If calibration fails multiple times, please promptly contact technical support and provide the error code.